ExperimentID | EXA000118 |
---|---|
ChemicalID | CH0000503 |
Chemical Name | Toxaphene 62 |
Official Method | No |
Chromatography | GC |
Chromatography mfr. | 7200 Q-TOF, Agilent |
Column | DB-35ms 30 m x 0.25 mm |
Column mfr. | Agilent |
Detector | TOF-MS |
Detector mfr. | 7200 Q-TOF, Agilent |
Ionization | NICI |
Extraction | SPE |
Extraction Details | Sequentially extracted with glass fiber filter GC50 and with SPE disk C18FF. GC50 filter residue: Ultrasonic extraction with acetone and Toluene.; Samples captured by C18FF: Eluted with Methanol, Acetone and Toluene |
Clean-up 1st | Washing |
Clean-up 1st Details | Washed in 5% sodium chloride aqueous solution |
Clean-up 2nd | Column cleanup |
Clean-up 2nd Details | Florisil, Sulfuric Acid/Silica Gel (50:50), Eluted with Dichloromethane/Hexane (20:80) |
Clean-up Spike | 13C12-PCB#3, #8, #11, #28, #331, #52, #77, #81, #101, #105, #114, #118, #123, #126, #138, #153, #156, #157, #167, #169, #170, #180, #189, #206, or #209 |
Syringe Spike | 13C12-PCB#9, #19, #70, #111, #155, #178, or #202 |
Derivatization | |
Recovery Correction | |
Instrument Detection Limit | |
Instrument Quantification Limit | |
Method Detection Limit | 20 pg/L |
Method Quantification Limit | 40 pg/L |
Repeatability | |
Standard | |
Recovery (CRM) | |
Recovery (spiked) | |
Recovery (IS) | |
Protocol | https://www.env.go.jp/chemi/kurohon/2019/shosai/3_7.pdf |
Remarks | Analysis method: p.437; Resolution: 10,000 |
RegisterDate | 2021/5/21 |
UpdateDate | 2021/5/21 |