| ExperimentID | EXA000146 |
|---|---|
| ChemicalID | CH0000215 |
| Chemical Name | PCP |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | 6890, Agilent / 7683, Agilent |
| Column | DB-5ms 30 m x 0.25 mm, 0.25 μm |
| Column mfr. | Agilent |
| Detector | HRMS |
| Detector mfr. | AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
| Ionization | SIM-EI |
| Extraction | LLE |
| Extraction Details | Extracted with Dichloromethane |
| Clean-up 1st | |
| Clean-up 1st Details | |
| Clean-up 2nd | Column cleanup |
| Clean-up 2nd Details | 5% Water-impregnated Silica Gel, Dichloromethane |
| Clean-up Spike | 13C12-PCP |
| Syringe Spike | 13C12-PCB#70 |
| Derivatization | BSTFA |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 9 pg/L |
| Method Quantification Limit | 24 pg/L |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2019/shosai/3_7.pdf |
| Remarks | Analysis method: p.450; Resolution: 10,000 |
| RegisterDate | 2021/5/21 |
| UpdateDate | 2021/5/21 |