ExperimentID |
EXA000150 |
ChemicalID |
CH0001180
|
Chemical Name |
ΣChlorinated tridecanes |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
7200 Q-TOF, Agilent |
Column |
DB-5ms 15 m x 0.25 mm |
Column mfr. |
Agilent |
Detector |
TOF-MS |
Detector mfr. |
7200 Q-TOF, Agilent |
Ionization |
EI or NICI |
Extraction |
SPE |
Extraction Details |
Sequentially extracted with glass fiber filter GC50 and with SPE disk C18FF. GC50 filter residue: Ultrasonic extraction with acetone and Toluene.; Samples captured by C18FF: Eluted with Methanol, Acetone and Toluene |
Clean-up 1st |
Washing |
Clean-up 1st Details |
Washed in 5% sodium chloride aqueous solution |
Clean-up 2nd |
Column cleanup |
Clean-up 2nd Details |
Florisil, Sulfuric Acid/Silica Gel (50:50), Eluted with Dichloromethane/Hexane (20:80) |
Clean-up Spike |
NA |
Syringe Spike |
13C12-PCB#111 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
1500 pg/L |
Method Quantification Limit |
4500 pg/L |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2019/shosai/3_7.pdf |
Remarks |
Analysis method: p.454 |
RegisterDate |
2021/5/21 |
UpdateDate |
2021/5/21 |