ExperimentID | EXA000214 |
---|---|
ChemicalID | CH0001178 |
Chemical Name | ΣChlorinated undecanes |
Official Method | No |
Chromatography | GC |
Chromatography mfr. | 7200 Q-TOF, Agilent |
Column | DB-5ms 15 m x 0.25 mm, 0.1 μm |
Column mfr. | Agilent |
Detector | TOF-MS |
Detector mfr. | 7200 Q-TOF, Agilent |
Ionization | EI or NICI |
Extraction | Solvent Extraction |
Extraction Details | Ultrasonic extraction: Extracted with Acetone. After filtration, wash the residue in Acetone.; Soxhlet Extraction: Extracted with Toluene. |
Clean-up 1st | Sulfate treatment |
Clean-up 1st Details | Sulfate |
Clean-up 2nd | Malti-layer column cleanup; GPC; Florisil column |
Clean-up 2nd Details | Malti-layer column cleanup: Silica Gel, Copper powder, Silica Gel Sulfate, Anhydrous sodium sulfate. Washed/Eluted with Dichloromethane/Hexane (20:80); Gel permeation chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC, Column: Shodex EV-2000 AC; Florisil column: Washed in Hexane. Eluted with Acetone. |
Clean-up Spike | 13C12-PCB#3, #8, #11, #28, #31, #52, #77, #81, #101, #105, #114, #118, #123, #126, #138, #153, #156, #157, #167, #169, #170, #180, #189, #206 or #209 |
Syringe Spike | 13C12-PCB#111 |
Derivatization | |
Recovery Correction | |
Instrument Detection Limit | |
Instrument Quantification Limit | |
Method Detection Limit | 5000 pg/g dry |
Method Quantification Limit | 15000 pg/g dry |
Repeatability | |
Standard | |
Recovery (CRM) | |
Recovery (spiked) | |
Recovery (IS) | |
Protocol | https://www.env.go.jp/chemi/kurohon/2019/shosai/3_7.pdf |
Remarks | Analysis method: p.438; Resolution: 13,000 |
RegisterDate | 2021/5/24 |
UpdateDate | 2021/5/24 |