| ExperimentID | EXA000277 |
|---|---|
| ChemicalID | CH0001174 |
| Chemical Name | Hexachloro-1,3-butadiene |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | HP 6890, Agilent |
| Column | DB-5ms 30 m x 0.25 mm, 0.25 μm |
| Column mfr. | Agilent |
| Detector | MS |
| Detector mfr. | HP 5973, Agilent |
| Ionization | SIM-EI |
| Extraction | Thermal Desorption (TD) |
| Extraction Details | Collection Tube for TD: Tenax TA |
| Clean-up 1st | |
| Clean-up 1st Details | |
| Clean-up 2nd | |
| Clean-up 2nd Details | |
| Clean-up Spike | |
| Syringe Spike | 13C6-Pentachlorobenzene |
| Derivatization | |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 10 pg/m3 |
| Method Quantification Limit | 30 pg/m3 |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2019/shosai/3_7.pdf |
| Remarks | Analysis method: p.450 |
| RegisterDate | 2021/6/04 |
| UpdateDate | 2021/6/04 |