| ExperimentID | EXA000298 |
|---|---|
| ChemicalID | CH0001618 |
| Chemical Name | Cyanamide |
| Official Method | No |
| Chromatography | LC |
| Chromatography mfr. | Prominence, Shimadzu |
| Column | InertSustain C18 150mm x 2.1mm, 5μm |
| Column mfr. | GL Sciences |
| Detector | MS/MS |
| Detector mfr. | 3200 QTRAP, AB Sciex |
| Ionization | SRM-ESI-negative |
| Extraction | LLE; SPE |
| Extraction Details | Extracted with Dichloromethane 3mL, Shaken then kept still standing; SPE extraction with Sep-Pak Silka Plus Long Cartridge 690 mg; Eluted with Dichloromethane/Acetonitrile (60:40) 10mL |
| Clean-up 1st | Filtration |
| Clean-up 1st Details | Filtered with GF/C filter 47mm, Whatman |
| Clean-up 2nd | Washing |
| Clean-up 2nd Details | Cartridge: Washed in Dichloromethane twice |
| Clean-up Spike | |
| Syringe Spike | |
| Derivatization | 10% Dansyl Chloride Acetone solution 60mg/mL and 4-Dimethylaminopyridine Acetone solution 200μL; Shaken and kept at 50 ℃ for 60 min. |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 280 ng/L |
| Method Quantification Limit | |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2020/shosai/5_1_2.pdf |
| Remarks | Analysis method: p.3 |
| RegisterDate | 2022/1/25 |
| UpdateDate | 2022/1/25 |