| ExperimentID | EXA000309 |
|---|---|
| ChemicalID | CH0001627 |
| Chemical Name | Valproic acid |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | TRACE GU Ultra, Thermo Fisher Scientific; 7890B, Agilent |
| Column | J&W DB-5ms 30m x 0.25mm, 0.25μm |
| Column mfr. | Agilent |
| Detector | MS |
| Detector mfr. | TSQ Quantum, Thermo Fisher Scientific; JMS-Q1500, JEOL |
| Ionization | SIM-EI |
| Extraction | SPE |
| Extraction Details | pH adjustment: pH 3; Sep-Pak Plus PS-2 10mL/min.; Eluted with Dichloromethane |
| Clean-up 1st | Washing |
| Clean-up 1st Details | Cartridge: Washed with purified water |
| Clean-up 2nd | Column cleanup |
| Clean-up 2nd Details | Supelclean LC-Si; Dichloromethane/Hexane (30:70) 10mL; Eluted with Dichloromethane/hexane (70:30) 10mL |
| Clean-up Spike | |
| Syringe Spike | Naphthalene-d8 |
| Derivatization | N-Methyl-N-tert-(butyldimethylsilyl)trifluoroacetamide |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 3.1 ng/L |
| Method Quantification Limit | |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2020/shosai/5_1_2.pdf |
| Remarks | Analysis method: p.8 |
| RegisterDate | 2022/1/25 |
| UpdateDate | 2022/1/25 |