| ExperimentID | EXA000362 |
|---|---|
| ChemicalID | CH0000144 |
| Chemical Name | βHCH |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | 6890, Agilent; 7683, Agilent |
| Column | HT8-PCB 60 m x 0.25 mm |
| Column mfr. | GL Sciences |
| Detector | HRMS |
| Detector mfr. | AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
| Ionization | SIM-EI |
| Extraction | SPE |
| Extraction Details | Sequentially extracted with glass fiber filter GC50 and with SPE disk C18FF; Ultrasonic extraction with Acetone; |
| Clean-up 1st | Multilayer column cleanup; GPC |
| Clean-up 1st Details | Multilayer column cleanup: Silica Gel, Silica Gel Sulfate(50:50), Anhydrous sodium sulfate; Eluted with Hexane (First fraction), Dichloromethane/Hexane (20:80) (Second fraction) |
| Clean-up 2nd | Column cleanup |
| Clean-up 2nd Details | Florisil; Eluted with Dichloromethane/Hexane (20:80) |
| Clean-up Spike | 13C6-β-HCH |
| Syringe Spike | 13C12-PCB#9, #19, #70, #111, #155, #178, or #202 |
| Derivatization | |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 1 pg/L |
| Method Quantification Limit | 3 pg/L |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
| Remarks | Analysis method: p.1; Resolution: 10,000 |
| RegisterDate | 2022/6/02 |
| UpdateDate | 2022/6/02 |