ExperimentID |
EXA000367 |
ChemicalID |
CH0000100
|
Chemical Name |
BDE47 |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
6890, Agilent; 7683, Agilent |
Column |
HT8-PCB 60 m x 0.25 mm |
Column mfr. |
GL Sciences |
Detector |
HRMS |
Detector mfr. |
AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
Ionization |
SIM-EI |
Extraction |
SPE |
Extraction Details |
Sequentially extracted with glass fiber filter GC50 and with SPE disk C18FF; Ultrasonic extraction with Acetone; |
Clean-up 1st |
Multilayer column cleanup; GPC |
Clean-up 1st Details |
Multilayer column cleanup: Silica Gel, Silica Gel Sulfate(50:50), Anhydrous sodium sulfate; Eluted with Hexane (First fraction), Dichloromethane/Hexane (20:80) (Second fraction) |
Clean-up 2nd |
Column cleanup |
Clean-up 2nd Details |
Florisil; Eluted with Dichloromethane/Hexane (20:80) |
Clean-up Spike |
13C12-PBDE#3, #15, #28, #47, #99, #126, #153, #154, #183, #197, #207 or #209 |
Syringe Spike |
PBDE#138; PBDE#206 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
4 pg/L |
Method Quantification Limit |
11 pg/L |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
Remarks |
Analysis method: p.1; Resolution: 10,000 |
RegisterDate |
2022/6/02 |
UpdateDate |
2022/6/02 |