ExperimentID |
EXA000431 |
ChemicalID |
CH0000827
|
Chemical Name |
Σhexa-BDEs |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
6890, Agilent; 7683, Agilent |
Column |
BPX-DXN 30 m x 0.25 mm or BP-1 15 m x 0.25 mm, 0.1 μm |
Column mfr. |
Trajan; Trajan |
Detector |
HRMS |
Detector mfr. |
AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
Ionization |
SIM-EI |
Extraction |
Solvent Extraction |
Extraction Details |
Ultrasonic extraction: Extracted with Acetone. After filtration, wash the residue in Acetone.; Soxhlet Extraction: Extracted with Toluene. |
Clean-up 1st |
Malti-layer column cleanup |
Clean-up 1st Details |
Malti-layer column cleanup: Silica Gel, Copper powder, Silica Gel Sulfate, Anhydrous sodium sulfate; Eluted with Hexane (First fraction), Dichloromethane/Hexane (20:80) (Second fraction) |
Clean-up 2nd |
GPC; Multi-layer column cleanup |
Clean-up 2nd Details |
Gel permeation chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC; Column: Shodex EV-2000 AC; Multi-layer column cleanup: Active carbon-dispersed silica gel, Sodium sulfate; Eluted with Dichloromethane/Hexane (25:75) |
Clean-up Spike |
13C12-PBDE#3, #15, #28, #47, #99, #126, #153, #154, #183, #197, #207 or #209 |
Syringe Spike |
PBDE#138; PBDE#206 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
2 pg/g dry |
Method Quantification Limit |
4 pg/g dry |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
Remarks |
Analysis method: p2-3; Resolution: 10,000 |
RegisterDate |
2022/7/12 |
UpdateDate |
2022/7/12 |