| ExperimentID | EXA000436 |
|---|---|
| ChemicalID | CH0000830 |
| Chemical Name | Σocta-BDEs |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | 6890, Agilent; 7683, Agilent |
| Column | BPX-DXN 30 m x 0.25 mm or BP-1 15 m x 0.25 mm, 0.1 μm |
| Column mfr. | Trajan; Trajan |
| Detector | HRMS |
| Detector mfr. | AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
| Ionization | SIM-EI |
| Extraction | Solvent Extraction |
| Extraction Details | Ultrasonic extraction: Extracted with Acetone. After filtration, wash the residue in Acetone.; Soxhlet Extraction: Extracted with Toluene. |
| Clean-up 1st | Malti-layer column cleanup |
| Clean-up 1st Details | Malti-layer column cleanup: Silica Gel, Copper powder, Silica Gel Sulfate, Anhydrous sodium sulfate; Eluted with Hexane (First fraction), Dichloromethane/Hexane (20:80) (Second fraction) |
| Clean-up 2nd | GPC; Multi-layer column cleanup |
| Clean-up 2nd Details | Gel permeation chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC; Column: Shodex EV-2000 AC; Multi-layer column cleanup: Active carbon-dispersed silica gel, Sodium sulfate; Eluted with Dichloromethane/Hexane (25:75) |
| Clean-up Spike | 13C12-PBDE#3, #15, #28, #47, #99, #126, #153, #154, #183, #197, #207 or #209 |
| Syringe Spike | PBDE#138; PBDE#206 |
| Derivatization | |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 1 pg/g dry |
| Method Quantification Limit | 3 pg/g dry |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
| Remarks | Analysis method: p2-3; Resolution: 10,000 |
| RegisterDate | 2022/7/12 |
| UpdateDate | 2022/7/12 |