ExperimentID |
EXA000442 |
ChemicalID |
CH0000933
|
Chemical Name |
mono-PCNs |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
6890, Agilent; 7683, Agilent |
Column |
DB-5ms 60 m x 0.32 mm |
Column mfr. |
Agilent |
Detector |
HRMS |
Detector mfr. |
AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
Ionization |
SIM-EI |
Extraction |
Solvent Extraction |
Extraction Details |
Extracted with Acetone. |
Clean-up 1st |
Malti-layer column cleanup |
Clean-up 1st Details |
Malti-layer column cleanup: Florisil, Anhydrous sodium sulfate; Eluted with Hexane (First fraction), Dichloromethane/Hexane (50:50) (Second fraction) |
Clean-up 2nd |
Malti-layer column cleanup |
Clean-up 2nd Details |
Malti-layer column cleanup: Silica Gel, Copper powder, Silica Gel Sulfate, Anhydrous sodium sulfate. Eluted with Hexane |
Clean-up Spike |
13C102-Chloronaphthalene |
Syringe Spike |
13C12-PCB#9, #19, #70, #111, #155, #178 or #202 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
0.9 pg/g dry |
Method Quantification Limit |
2.3 pg/g dry |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
Remarks |
Analysis method: p4; Resolution: 10,000 |
RegisterDate |
2022/7/12 |
UpdateDate |
2022/7/12 |