ExperimentID |
EXA000456 |
ChemicalID |
CH0001520
|
Chemical Name |
Dicofol |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
6890, Agilent; 7683, Agilent |
Column |
DB-5ms 15 m x 0.25 mm |
Column mfr. |
Agilent |
Detector |
HRMS |
Detector mfr. |
AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
Ionization |
SIM-EI |
Extraction |
Solvent Extraction |
Extraction Details |
Extracted with Acetone |
Clean-up 1st |
Washing |
Clean-up 1st Details |
Washed in 5% Sodium chloride aqueous solution |
Clean-up 2nd |
Multilayer column cleanup; GPC |
Clean-up 2nd Details |
Multilayer column cleanup: Frorigil, Anhydrous sodium sulfate, Eluted with Hexane (First fraction) and Dichloromethane/Hexane(50:50)(Second fraction); Only second fraction; Gel Permeation Chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC; Column: Shodex EV-2000 AC |
Clean-up Spike |
Spiked, but the details are not available |
Syringe Spike |
13C12-PCB#9, #19, #70, #111, #155, #178, or #202 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
2 pg/g dry |
Method Quantification Limit |
4 pg/g dry |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2020/shosai/5_3_2.pdf |
Remarks |
Analysis method: p23; Resolution: 10,000 |
RegisterDate |
2022/7/12 |
UpdateDate |
2022/7/12 |