ExperimentID |
EXA000592 |
ChemicalID |
CH0001655
|
Chemical Name |
Hexachloroethane |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
[1] 6890N, Agilent; [2] 8890, Agilent |
Column |
HP-5ms UI 30 m x 0.25 mm, 0.25 μm |
Column mfr. |
Agilent |
Detector |
HRMS |
Detector mfr. |
[1] 5975C, Agilent; [2] JMS-Q1500, JEOL |
Ionization |
SIM-EI |
Extraction |
Solvent extraction |
Extraction Details |
Extracted with Sodium chloride (Fresh water only), Dichloromethane, shaken twice |
Clean-up 1st |
Column Cleanup |
Clean-up 1st Details |
LC-Si Glass Tubes, Eluted with Hexane |
Clean-up 2nd |
|
Clean-up 2nd Details |
|
Clean-up Spike |
|
Syringe Spike |
naphthalene-d8; p-Bromofluorobenzene |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
0.55 ng/L |
Method Quantification Limit |
|
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_1_2.pdf |
Remarks |
Analysis method: p.2 |
RegisterDate |
2022/10/17 |
UpdateDate |
2022/10/17 |