ExperimentID |
EXA000599 |
ChemicalID |
CH0001660
|
Chemical Name |
Triglycidyl isocyanurate |
Official Method |
No |
Chromatography |
LC |
Chromatography mfr. |
ExionLC, AB Sciex |
Column |
CORTECS C18+ 150 mm x 2.1 mm, 2.7 μm |
Column mfr. |
Waters |
Detector |
MS/MS |
Detector mfr. |
Triple Quad 4500 System, AB Sciex |
Ionization |
SRM-ESI-positive |
Extraction |
|
Extraction Details |
Air collection: Sep-Pak Silica Plus Long Cartridge, 690 mg 0.5L/min. x 24h.; Eluted with Methanol |
Clean-up 1st |
|
Clean-up 1st Details |
|
Clean-up 2nd |
|
Clean-up 2nd Details |
|
Clean-up Spike |
|
Syringe Spike |
|
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
0.039 ng/m3 |
Method Quantification Limit |
|
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_1_2.pdf |
Remarks |
Analysis method: p.4 |
RegisterDate |
2022/10/17 |
UpdateDate |
2022/10/17 |