ExperimentID |
EXA000601 |
ChemicalID |
CH0000907
|
Chemical Name |
Aniline |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
8890, Agilent |
Column |
J&W DB-WAX 30 m x 0.25 mm, 0.25 μm |
Column mfr. |
Agilent |
Detector |
MS |
Detector mfr. |
5977B, Agilent |
Ionization |
SIM-EI |
Extraction |
SPE |
Extraction Details |
Sep-Pak PS-2 Plus; Eluted with Methyl acetate |
Clean-up 1st |
Washing |
Clean-up 1st Details |
Washed in purified water |
Clean-up 2nd |
|
Clean-up 2nd Details |
|
Clean-up Spike |
Aniline-d5 |
Syringe Spike |
Naphthalene-d8 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
14 ng/L |
Method Quantification Limit |
|
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_2_2.pdf |
Remarks |
Analysis method: p.1 |
RegisterDate |
2022/11/25 |
UpdateDate |
2022/11/25 |