ExperimentID |
EXA000627 |
ChemicalID |
CH0001666
|
Chemical Name |
[(3-Hexadecanamidopropyl)(dimethyl)ammonio]acetate |
Official Method |
No |
Chromatography |
LC |
Chromatography mfr. |
1260, Agilent |
Column |
Inertsil ODS-4 HP 150 mm x 2.1 mm, 3 μm |
Column mfr. |
GL Sciences |
Detector |
MS/MS |
Detector mfr. |
QTRAP 4500QT, AB Sciex |
Ionization |
SRM-ESI-positive |
Extraction |
Solvent Extraction |
Extraction Details |
Extracted with 28 % Ammonia water / Methanol (5:95) 30 mL, Shaken and kept still standing at 70 ℃ for 15 min. Then, centrifuged 3,000 rpm for 10 min. ( 20 min. in the 3rd round); Repeated these procedures three times |
Clean-up 1st |
Column cleanup |
Clean-up 1st Details |
Sep-Pak PS2 Plus Short Cartridge, 225 mg / 80 μm; Run the extract through the cartridge at 2 mL / min.; Remove interfering substance with purified water 2 mL, formic acid / purified water (2:98) 5 mL, Methanol 10 mL; Eluted with 28 % Ammonia water / Methanol (5:95) 10 mL. |
Clean-up 2nd |
Column cleanup |
Clean-up 2nd Details |
Oasis MCX Plus Short Cartridge 225 mg / 60 μm; Run the extract (through the former column cleanup) through the cartridge at 2 mL / min.; Remove interfering substance with purified water 5 mL; Eluted with Methanol 10 mL. |
Clean-up Spike |
|
Syringe Spike |
|
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
0.19 ng/g dry |
Method Quantification Limit |
|
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_2_2.pdf |
Remarks |
Analysis method: p.2 |
RegisterDate |
2022/12/08 |
UpdateDate |
2022/12/08 |