ExperimentID |
EXA000753 |
ChemicalID |
CH0001174
|
Chemical Name |
Hexachloro-1,3-butadiene |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
7890, Agilent |
Column |
DB-5ms 30 m x 0.25 mm, 0.25 μm |
Column mfr. |
Agilent |
Detector |
MS/MS |
Detector mfr. |
7010, Agilent |
Ionization |
MRM-EI |
Extraction |
Reflux extraction |
Extraction Details |
Extracted with Hexane; Reflux boiling for 60 min.; Dehydration with Whatman 1PS filter |
Clean-up 1st |
|
Clean-up 1st Details |
|
Clean-up 2nd |
|
Clean-up 2nd Details |
|
Clean-up Spike |
13C4-Hexachloro-1,3-butadiene |
Syringe Spike |
Naphthalene-d8; Acenaphtene-d10; Phenanthrene-d10; Fluoranthene-d10; Chrysene-d12 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
40 pg/L |
Method Quantification Limit |
100 pg/L |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_3_2.pdf |
Remarks |
Analysis method: p.9 |
RegisterDate |
2023/1/26 |
UpdateDate |
2023/1/26 |