ExperimentID |
EXA000754 |
ChemicalID |
CH0001177
|
Chemical Name |
ΣChlorinated decanes |
Official Method |
No |
Chromatography |
GC |
Chromatography mfr. |
6890, Agilent; 7890, Agilent |
Column |
DB-5ms 15 m x 0.25 mm |
Column mfr. |
Agilent |
Detector |
TOF-MS |
Detector mfr. |
AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
Ionization |
EI or NICI |
Extraction |
SPE |
Extraction Details |
Sequentially extracted with glass fiber filter GC50 and with SPE disk C18FF; Ultrasonic extraction with Acetone 50 mL, 20 min. x 3 times |
Clean-up 1st |
Column cleanup |
Clean-up 1st Details |
Silica gel, Silica gel sulfate, Anhydrous sodium sulfate; Eluted with Dichloromethane/Hexane (20:80) (Second fraction) |
Clean-up 2nd |
Column cleanup |
Clean-up 2nd Details |
Florisil, Anhydrous sodium sulfate; Eluted with Dichloromethane/Hexane (20:80) |
Clean-up Spike |
1,5,5,6,6,10-Hexachlorinated decanes-13C10 |
Syringe Spike |
Naphthalene-d8; Acenaphtene-d10; Phenanthrene-d10; Fluoranthene-d10; Chrysene-d12 |
Derivatization |
|
Recovery Correction |
|
Instrument Detection Limit |
|
Instrument Quantification Limit |
|
Method Detection Limit |
200 pg/L |
Method Quantification Limit |
400 pg/L |
Repeatability |
|
Standard |
|
Recovery (CRM) |
|
Recovery (spiked) |
|
Recovery (IS) |
|
Protocol |
https://www.env.go.jp/chemi/kurohon/2021/shosai/5_3_2.pdf |
Remarks |
Analysis method: p.1-2; Resolution: 10,000 |
RegisterDate |
2023/1/26 |
UpdateDate |
2023/1/26 |