| ExperimentID | EXA000776 |
|---|---|
| ChemicalID | CH0000074 |
| Chemical Name | PCB167 |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | 6890, Agilent; 7890, Agilent |
| Column | HT8-PCB 60 m x 0.25 mm |
| Column mfr. | GL Sciences |
| Detector | HRMS |
| Detector mfr. | AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
| Ionization | SIM-EI |
| Extraction | Solvent Extraction |
| Extraction Details | Extracted with Acetone; Shaken for 10 min., Sonicated for 15min., then Centrifugated (3,000rpm, 10 min.); Repeated these three times |
| Clean-up 1st | Column cleanup |
| Clean-up 1st Details | Silica Gel, Silica Gel Sulfate, Anhydrous sodium sulfate, Copper powder; Eluted with Hexane (First fraction), Dichloromethane/Hexane (20:80) (Second fraction) |
| Clean-up 2nd | GPC |
| Clean-up 2nd Details | Gel permeation chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC; Column: Shodex EV-2000 AC |
| Clean-up Spike | 13C12-PCB#3, #8, #11, #28, #31, #52, #77, #81, #101, #105, #114, #118, #123, #126, #138, #153, #156, #157, #167, #169, #170, #180, #189, #206 or #209 |
| Syringe Spike | 13C12-PCB#9, #19, #70, #111, #155, #178, #202 |
| Derivatization | |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 0.3 pg/g dry |
| Method Quantification Limit | 0.8 pg/g dry |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2021/shosai/5_3_2.pdf |
| Remarks | Analysis method: p.3; Resolution: 10,000 |
| RegisterDate | 2023/2/15 |
| UpdateDate | 2023/2/15 |