| ExperimentID | EXA000792 |
|---|---|
| ChemicalID | CH0001172 |
| Chemical Name | cis-Heptachlor epoxide |
| Official Method | No |
| Chromatography | GC |
| Chromatography mfr. | 6890, Agilent; 7890, Agilent |
| Column | RH-12ms 30 m x 0.25 mm |
| Column mfr. | INVENTX |
| Detector | HRMS |
| Detector mfr. | AutoSpec Ultima, Waters; AutoSpec Premier, Waters |
| Ionization | SIM-EI |
| Extraction | Solvent Extraction |
| Extraction Details | Extracted with Acetone; Shaken for 10 min., Sonicated for 15min., then Centrifugated (3,000rpm, 10 min.); Repeated these three times |
| Clean-up 1st | Column cleanup |
| Clean-up 1st Details | Florisil, Anhydrous sodium sulfate; Remove interfering substance with Hexane; Eluted with Dichloromethane/Hexane (20:80) |
| Clean-up 2nd | GPC |
| Clean-up 2nd Details | Gel permeation chromatography: Acetone/Cyclohexane (20:80), Precolumn: Shodex EV-G AC; Column: Shodex EV-2000 AC |
| Clean-up Spike | 13C10-cis-heptachlor epoxide |
| Syringe Spike | 13C12-PCB#9, #19, #70, #111, #155, #178, #202 |
| Derivatization | |
| Recovery Correction | |
| Instrument Detection Limit | |
| Instrument Quantification Limit | |
| Method Detection Limit | 0.7 pg/g dry |
| Method Quantification Limit | 1.7 pg/g dry |
| Repeatability | |
| Standard | |
| Recovery (CRM) | |
| Recovery (spiked) | |
| Recovery (IS) | |
| Protocol | https://www.env.go.jp/chemi/kurohon/2021/shosai/5_3_2.pdf |
| Remarks | Analysis method: p.3-4; Resolution: 10,000 |
| RegisterDate | 2023/2/15 |
| UpdateDate | 2023/2/15 |